MANUAL (RED LED LIGHT MK-02GX)

Transilluminator Vein Finder for neonatal & infant use

Red LED Light MK-02GX, Transilluminator Vein Finder for neonatal & infant use
            User's Manual

You are kindly requested to use the Red LED Light MK-02GX proprely as explained below.


General description
Red LED Light uses high power red LED which emits light of wavelength close to the "optical window",
having deeper light penetration into the human tissue and radiates far less heat than the black body
radiation filament lamp.
Thanks further to the use of PWM technology and higher thermally conductive & high temperature
resistive resin, the heat is well managed to be designed friendly to the skin of the neonate & infant.
Red LED Light is a tiny but strong ellipsoidal vein finder with excellent palm fitness feeling for helping
you to pinpoint the optimal vein area for IV insertion


Specifications

Battery:2 x AA alkaline battery
Material:Noryl, PO/PPE+PS(lamp housing); PC(light window)/PC+ABS(power unit)/
Soft, slim & highly flexible robot cable(power cord)
Temperature resistance of the lighting unit:will not deflect below 130deg.,C
Chemical resistance of the lighting unit:high resistance to many organic & inorganic chemicals
Weight:lighting unit/4.8gr;power unit/60gr(2xAA not_included);cord/7gr
Size:lighting unit/30(L)x20(W)x14(H)mm;power unit/84(L)x53(W)x25(H)mm;cord/80cm(L)xФ2.4mm
Disinfection:It can be disinfected by ethanol dipped sterile cotton before & after use


Prohibited actions
★Do not stare at the light source while the power is turned on. In addition, never direct the light towards
 other people's eyes.
★Avoid its long & continuous operation as it may burn the palm skin.
★Do not pour liquid neither on the lighting nor the power units.
★Always replace with new 2xAA alkaline battery
★Never pull the cord strongly or wave the lighting unit by holding the cord.


How to operate Red LED Light

1. Loading & Replacing the battery
   As no battery is loaded on shipment from the factory, kindly first load the battery case with new
   2xAA battery. When you replace them, first confirm that the switch is set to its Off position.
   Then, pull-up the lever to open the lid and load 2 x AA battery in the correct direction
   as depicted. Then snap back the lid to latch.

2.Method of operation
  A. Place the light window in direct contact with the area you want to trans-illuminate.
  B. Turn on the power switch.
  C. When the light becomes unnecessary or doctor feels skin temperature of the patient has risen rather
high, it should be put out immediately.
The integral time of the most recent irradiating operations on a particular area should not exceed 3 minutes
to avoid any thermal injuries.




It is ideally suitable to trans-illuminate the tiny vein of the neonate.

Cautions

● Do not disassemble nor revamp it.
● Do not shock nor throw it.
● Do not bind the battery case with the cord tightly.
● Although the light output is always maintained at the same level, the battery should be changed
  whenever the light output becomes either unstable or insufficient.

Warranty

The device is warranted for one year from the date of your purchase.
For warranty, the proof of your purchase date and serial number are required.

name LED transilluminator
serial # indicated inside the power unit
FDA registration # 3006737183
operator # 9097725
listing # E579683

manufacturer & distributor
name
esu llp
address
2-2-7 Akanedai Aoba Yokohama JAPAN 227-0066
phone
+81 45 512 9560